Why?
There are mainly two types of LC measuring methods: Touch Current (TC) and Protective Conductor Current (PCC). Touch Current is the current that flows when a human body touches
the equipment. Protective Conductor Current is the current that flows from the unit through the grounding conductor into a household ground. LC is capable of performing tests that comply with standards for a wide range of equipment.
LC is capable of performing tests that comply with standards for a wide range of equipment.
Dut input power capacity 300 VAC / 16A max.
Supports all touch current and protective conductor current (earth leakage current) tests.
Quanti Leakage Current test is designed to perform leakage current (touch current and protective conductor current) tests on general electric equipment (medical option on request). It enables you to conduct tests that conform the requirements of the applicable IEC, UL, EN and other standards.
Power supply separation
The power supply for the Quanti Leakage Current is separate from the line power supply of the equipment that is being tested, preventing this way any damage to equipment due to wrong supply voltage being in the input.
Conforms to following standards among others:
UL 1563 - Standard for electric spas, equipment assemblies and associated equipment
UL/IEC/EN 61010-1 - Electrical equipment for measurement, control and laboratory use
UL/IEC/EN 60950-1 - Information technology equipment
UL/IEC/EN 60335-1 - Safety of household and similar electrical appliances
IEC/EN 60990 - Methods of measurement of touch current and protective conductor current
Measurement modes
LEAKAGE CURRENT BETWEEN ENCLOSURE AND LINE
LEAKAGE CURRENT BETWEEN ENCLOSURE AND EARTH
LEAKAGE CURRENT BEWEEN ENCLOSURE AND ENCLOSURE
EARTH LEAKAGE CURRENT
Touch current (TC and protective conductor current (PCC) test
Eight measurement circuit networks conforming to the applicable standards are provided as standard. The switching of the polarities of the power line to the DUT, as well as single-fault conditions, are automatically set with relays inside the tester.
Leakage current function specifications
| DUT input power capacity |
>AC: 300V / 16A max., separate input connection | |
| Short protection |
External |
|
| Current range AC/DC |
50.0µA - 25mA (75mA peak) |
|
| Line voltage measurement |
0 - 300 VAC |
|
| Line current measurement |
0 - 16A |
|
| Timer function |
0; 0.3 - 999.9sec (0 = continuous) |
|
| Line conditions |
Reverse power switch: Reverse polarity switch for normal condition Neutral switch: Neutral switch on/off selection for single fault Ground switch: Ground switch on/off selection for class 1 single fault |
|
| HI and LO limit |
Adjustable within the full measuring range |